Preliminary evaluation of decontamination with a 10-voltsupplied plasma jet: made possible by piezoelectric generator - Archive ouverte HAL
Article Dans Une Revue IEEE Transactions on Plasma Science Année : 2020

Preliminary evaluation of decontamination with a 10-voltsupplied plasma jet: made possible by piezoelectric generator

Résumé

This paper deals with preliminary decontamination performances obtained from a plasma jet generated by a piezoelectric transformer. This low voltage supply solution (<10V) is investigated as a compact and safe solution for the decontamination of medical thermo-sensitive devices. Promising reduction rates are obtained on Pseudomonas aeruginosa and Staphylococcus aureus bacteria with an argon plasma jet at atmospheric pressure about 2.5W input power.
Fichier principal
Vignette du fichier
TPS12256_Proof_hi.pdf (1.36 Mo) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-02640823 , version 2 (28-05-2020)
hal-02640823 , version 1 (19-07-2022)

Identifiants

Citer

Jordan Stekke, Claire Tendero, Pascal Tierce, Christian Courtois, Gaetan Rauwel, et al.. Preliminary evaluation of decontamination with a 10-voltsupplied plasma jet: made possible by piezoelectric generator. IEEE Transactions on Plasma Science, 2020, 48 (5), pp.1264-1270. ⟨10.1109/TPS.2020.2985406⟩. ⟨hal-02640823v1⟩