Mass flow rate and permeability measurements in microporous media - Archive ouverte HAL Access content directly
Journal Articles Vacuum Year : 2018

Mass flow rate and permeability measurements in microporous media

Abstract

10 The transient method of the mass flow rate measurements through a microp-orous media is developed and analyzed. This method is based on the constant volume technique and the exponential fit of the pressure evolution in each tank which allows calculating the permeability directly. The pressure relaxation time, a single fitting parameter, is introduced and its behaviors are analyzed in a large pressure range. By measuring the pressure relaxation time for one gas, the per-meability of a microporous sample can be derived for the other gases. With the actual experimental setup, we measured the mass flow rate through the micro-porous media in the range 5 · 10 −7 − 5 · 10 −12 [kg s −1 ] and the permeability in the range 10 −14 − 10 −11 [m 2 ].
Fichier principal
Vignette du fichier
perm_microporous_02.pdf (761.49 Ko) Télécharger le fichier
Origin : Files produced by the author(s)
Loading...

Dates and versions

hal-01888007 , version 1 (08-03-2019)

Identifiers

Cite

Martin Victor Johansson, Fabrice Testa, Imen Zaier, Pierre Perrier, Jean Philippe Bonnet, et al.. Mass flow rate and permeability measurements in microporous media. Vacuum, 2018, 158, pp.75-85. ⟨10.1016/j.vacuum.2018.09.030⟩. ⟨hal-01888007⟩
107 View
714 Download

Altmetric

Share

Gmail Facebook Twitter LinkedIn More