Résumé

This paper proposes a new solution to the problem of torque minimization of the medical device SurgiScope ® by connecting to the initial structure a secondary mechanical system, which generates a vertical constant force on the platform of the robotized device. The conditions for optimization are formulated by the minimization of the root-mean-square values of the input torques of the studied device. The positioning errors of the unbalanced and balanced robots are provided. A significant reduction of these errors is achieved by using the suggested balancing mechanism. The efficiency of the developed approach is illustrated by numerical simulations.
Fichier principal
Vignette du fichier
Delta_ASME_V04_HAL.pdf (1.18 Mo) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-00451933 , version 1 (26-06-2019)

Identifiants

Citer

Sébastien Briot, Cédric Baradat, Sylvain Guegan, Vigen Arakelian. Contribution to the Improvement of the Medical Device SurgiScope. 2007 ASME International Design Engineering Technical Conferences (DETC 2007), Sep 2007, Las Vegas, NV, United States. pp.653-661, ⟨10.1115/DETC2007-35067⟩. ⟨hal-00451933⟩