Assessment of guide fitting using an intra-oral scanner: An in vitro study

Résumé

Objectives: The aim of this study was to evaluate the capacity of an intra-oral scanner (IOS) to assess the position of an endodontic guide in vitro.Methods: Fourteen extracted human teeth were placed into a maxillary model and scanned using computed tomography and a reference laboratory scanner. An ideal endodontic guide was then created and modified by adding defects of different thicknesses to simulate incorrect positions: 50 & mu;m, 150 & mu;m, 400 & mu;m, and 1000 & mu;m. For each thickness, guides were printed three times and each guide was scanned by three experimented operators using a Trios 4 IOS (3Shape, Copenhagen, Denmark). The 36 scans were compared using a best-fit alignment to the master model without defect to define the accuracy of the method and the positioning error.Results: The IOS presented a mean trueness of 1.28 & mu;m (SD= 12.70) and a mean precision of 11.52 & mu;m (SD= 62.17). Considering all sizes of defect, the mean measured position of the endodontic guide was highly correlated (R>0.99) with the expected position. Compared to the ideal guide, there was a mean linear deviation of 46.11 & mu;m (SD= 23.21) and a mean angular deviation of 5.9 degrees (SD= 1.2); this deviation was not influenced by the operator.Conclusion: The present study found that an IOS had good performance to detect a positioning error of the endodontic guide in vitro.Clinical Significance: This new application of IOS has a promising potential in clinical practice to assist practitioners during the fitting of guides.

Fichier principal
Vignette du fichier
S0300571223001768.pdf (332.75 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Licence

Dates et versions

hal-04661816 , version 1 (09-07-2025)

Licence

Identifiants

Citer

Elias Bittar, Pauline Binvignat, Cyril Villat, Jean-Christophe Maurin, Maxime Ducret, et al.. Assessment of guide fitting using an intra-oral scanner: An in vitro study. Journal of Dentistry, 2023, 135, ⟨10.1016/j.jdent.2023.104590⟩. ⟨hal-04661816⟩
200 Consultations
170 Téléchargements

Altmetric

Partager

  • More