Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer

Résumé

Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V_{95%} to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of 1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors.
Fichier principal
Vignette du fichier
2024 Heikkilä EJMP.pdf (984.35 Ko) Télécharger le fichier
mmc1.pdf (766.9 Ko) Télécharger le fichier
Origine : Publication financée par une institution
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-04366710 , version 1 (02-01-2024)

Licence

Paternité

Identifiants

Citer

Annele Heikkilä, Eeva Boman, Maija Rossi, Antti Vanhanen, Mikko Mankinen, et al.. Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer. Physica Medica, 2024, 117, pp.103203. ⟨10.1016/j.ejmp.2023.103203⟩. ⟨hal-04366710⟩
72 Consultations
13 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More