Evaluation of stereotactic VMAT lung treatment plans for small moving targets

Résumé

Purpose: The aim of this study is to perform patient quality controls and end-to-end tests for stereotactic VMAT lung treatment plans and to investigate the influence of various parameters on the results. Method: 18 plans were defined by an experimental design methodology to cover a large variety of stereotactic VMAT lung treatments including different doses per fraction, target diameters, target movements and respiratory parameters. Plans were first controlled using portal dosimetry and a homogeneous static cylindrical phantom. End-to-end tests were then performed in a dynamic respiratory thorax phantom. Measurements were conducted with ionization chamber and films. Calculations were performed with the AcurosXB and AAA algorithms in 6 FFF. Results: Portal dosimetry gave excellent gamma pass rates (greater than 97.1 %) and dose deviations between measurement and calculations in a homogeneous static phantom were smaller than 2 %. The methodology followed for comparing calculated and measured doses in a moving target was validated in static fields (largest deviation smaller than 2 %). End-to-end tests showed mean deviations of 1.9 %, 3.3 % and 6.6 % for the 3, 2 and 1 cm diameter’s target respectively. Deviations increased for larger movements for the 1 cm lesion. Conclusion: End-to-end tests revealed that stereotactic VMAT lung treatment plans for moving targets can be delivered within 5 % for 3 and 2 cm diameter targets and amplitudes up to 1.5 cm. The AcurosXB and AAA algorithms however tend to underestimate the dose to the target. Even with satisfactory patient quality controls like portal dosimetry, extra care should be taken for GTV lesions smaller than 2 cm.
Fichier sous embargo
Fichier sous embargo
0 10 6
Année Mois Jours
Avant la publication
samedi 1 mars 2025
Fichier sous embargo
samedi 1 mars 2025
Connectez-vous pour demander l'accès au fichier

Dates et versions

hal-04077750 , version 1 (12-06-2023)

Identifiants

Citer

Stéphane Dufreneix, C. Kirié, D. Autret. Evaluation of stereotactic VMAT lung treatment plans for small moving targets. Physica Medica, 2023, 107, pp.102547. ⟨10.1016/j.ejmp.2023.102547⟩. ⟨hal-04077750⟩
25 Consultations
3 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More