Elimination of cefotaxime using polysulfone and polyacrylonitrile-derived filters: An in vitro assessment - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue The International Journal of Artificial Organs Année : 2023

Elimination of cefotaxime using polysulfone and polyacrylonitrile-derived filters: An in vitro assessment

Résumé

Continuous renal replacement therapy (CCRT) efficiently eliminates cefotaxime. To our knowledge, there are no previous in vitro studies dealing with the disposition of cefotaxime. We studied the elimination of cefotaxime by two filters in a model mimicking a session of CRRT using the NeckEpur ® technology. The ST150 ® -polyacrylonitrile filter with the Prismaflex, Baxter-Gambro, and the AV1000 ® -polysulfone filter with the Multifiltrate Pro, Fresenius, were studied. Continuous filtration used a flowrate of 1 L/h in post-dilution only. Simulated blood flowrate was set at 200 mL/min. Routes of elimination were assessed using the NeckEpur ® technology. Cefotaxime concentrations were measured using ultra high-performance liquid chromatography, and tandem mass spectrometry. Two sessions were performed using the ST ® filter and three using the AV ® filter. Stability of cefotaxime during 6 h was assessed in triplicate with a mean variation of concentrations of 2.4 ± 1.5% at the end of the study. The mean measured initial concentration in the central compartment (CC) for the five sessions was 52.4 mg/L. The mean amount eliminated from the CC at the end of the sessions using the ST150 ® -polyacrylonitrile and the AV1000 ® -polysulfone filters were 72% and 73%, respectively. The clearances of cefotaxime from the central compartment (CC) were 1.1 and 1.2 L/h, respectively. The mean sieving coefficient were 0.99 and 0.99, respectively. The mean percentages of the amount eliminated from the CC by filtration/adsorption were 87/13% and 92/8%, respectively. Both adsorption percentages were below 15%. We conclude neither the ST150 ® -polyacrylonitrile nor the AV1000 ® -polysulfone filters result in clinically significant adsorption of cefotaxime.
Fichier non déposé

Dates et versions

hal-04068124 , version 1 (13-04-2023)

Identifiants

Citer

Jessica Le Ven, Camille Pellan, Valentin Maulet, Alban Le Monnier, Frédéric Baud. Elimination of cefotaxime using polysulfone and polyacrylonitrile-derived filters: An in vitro assessment. The International Journal of Artificial Organs, 2023, 46 (2), pp 113-119. ⟨10.1177/03913988221143803⟩. ⟨hal-04068124⟩
60 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More