Résumé

Assessing cortisol levels in human bodies has become essential to diagnose heart failure (HF). In this work, we propose a salivary cortisol detection strategy as part of an easily integrable labon-a-chip for detection of HF biomarkers. Our developed capacitive immunosensor based on hafnium oxide (HfO 2)/silicon structure showed good linearity between increasing cortisol concentration and the charge-transfer resistance/capacitance. Moreover, the developed biosensor was demonstrated to be highly selective toward cortisol compared to other HF biomarkers such as tumor necrosis factor (TNF-α) and N-terminal pro-brain natriuretic peptide (NT-proBNP). The precision of our developed biosensor was evaluated, and the difference between the determined cortisol concentration in saliva and its expected one is < 18%.
Fichier principal
Vignette du fichier
micromachines-13-02235-v2.pdf (2.52 Mo) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte

Dates et versions

hal-03997539 , version 1 (20-02-2023)

Identifiants

Citer

Hamdi Ben Halima, Nadia Zine, Joan Bausells, Nicole Jaffrezic-Renault, Abdelhamid Errachid. A Novel Cortisol Immunosensor Based on a Hafnium Oxide/Silicon Structure for Heart Failure Diagnosis. Micromachines, 2022, 13 (12), pp.2235. ⟨10.3390/mi13122235⟩. ⟨hal-03997539⟩