Article Dans Une Revue Journal of Synchrotron Radiation Année : 2020

White beam diagnostics using X-ray back-scattering from a CVD diamond vacuum window

Résumé

Collecting back-scattered X-rays from vacuum windows using a pinhole X-ray camera provides an efficient and reliable method of measuring the beam shape and position of the white synchrotron beam. In this paper, measurements are presented that were conducted at ESRF beamline ID6 which uses an in-vacuum cryogenically cooled permanent-magnet undulator (CPMU18) and a traditional U32 undulator as its radiation sources, allowing tests to be performed at very high power density levels that were adjusted by changing the gap of the undulators. These measurements show that it is possible to record beam shape and beam position using a simple geometry without having to place any further items in the beam path. With this simple test setup it was possible to record the beam position with a root-mean-square noise figure of 150 nm.

Mots clés

Fichier principal
Vignette du fichier
J_Synchrotron_Rad_27_37.pdf (827.03 Ko) Télécharger le fichier
Origine Fichiers éditeurs autorisés sur une archive ouverte
Licence

Dates et versions

hal-03725793 , version 1 (18-07-2022)

Licence

Identifiants

Citer

Roelof van Silfhout, Daniel Pothin, Thierry Martin. White beam diagnostics using X-ray back-scattering from a CVD diamond vacuum window. Journal of Synchrotron Radiation, 2020, 27, pp.37-43. ⟨10.1107/S1600577519015340⟩. ⟨hal-03725793⟩

Collections

28 Consultations
83 Téléchargements

Altmetric

Partager

  • More