Impact of Spacing and Orientation on the Scar Threshold With a High-Density Grid Catheter

Masateru Takigawa
  • Fonction : Auteur
Jatin Relan
  • Fonction : Auteur
  • PersonId : 761727
  • IdRef : 167644211
Takeshi Kitamura
  • Fonction : Auteur
Claire A. Martin
  • Fonction : Auteur
Steven Kim
  • Fonction : Auteur
Ruairidh Martin
  • Fonction : Auteur
Ghassen Cheniti
  • Fonction : Auteur
Konstantinos Vlachos
  • Fonction : Auteur
Grégoire Massoullié
  • Fonction : Auteur
Antonio Frontera
  • Fonction : Auteur
Nathaniel Thompson
  • Fonction : Auteur
Michael Wolf
  • Fonction : Auteur
Felix Bourier
  • Fonction : Auteur
Anna Lam
  • Fonction : Auteur
Josselin Duchateau
  • Fonction : Auteur
Thomas Pambrun
  • Fonction : Auteur
Arnaud Denis
  • Fonction : Auteur
Nicolas Derval
  • Fonction : Auteur
Xavier Pillois
  • Fonction : Auteur
Julie Magat
Jerome Naulin
  • Fonction : Auteur
Mathilde Merle
  • Fonction : Auteur
Florent Collot
  • Fonction : Auteur
Hubert Cochet
  • Fonction : Auteur
  • PersonId : 880054
Mélèze Hocini
  • Fonction : Auteur
Michel Haïssaguerre
  • Fonction : Auteur
Frederic Sacher
  • Fonction : Auteur
Pierre Jaïs
  • Fonction : Auteur

Résumé

BACKGROUND: Multipolar catheters are increasingly used for highdensity mapping. However, the threshold to define scar areas has not been well described for each configuration. We sought to elucidate the impact of bipolar spacing and orientation on the optimal threshold to match magnetic resonance imaging-defined scar. METHOD: The HD-Grid catheter uniquely allows for different spatially stable bipolar configurations to be tested. We analyzed the electrograms with settings of HD-16 (3 mm spacing in both along and across bipoles) and HD-32 (1 mm spacing in along bipoles and 3 mm spacing in across bipoles) and determined the optimal cutoff for scar detection in 6 infarcted sheep. RESULTS: From 456 total acquisition sites (mean 76±12 per case), 14 750 points with the HD-16 and 32286 points with the HD-32 configuration for bipolar electrograms were analyzed. For bipolar voltages, the optimal cutoff value to detect the magnetic resonance imaging-defined scar based on the Youden's Index, and the area under the receiver operating characteristic curve (AUROC) differed depending on the spacing and orientation of bipoles; across 0.84 mV (AUROC, 0.920; 95% CI, 0.911-0.928), along 0.76 mV (AUROC, 0.903; 95% CI, 0.893-0.912), northeast direction 0.95 mV (AUROC, 0.923; 95% CI, 0.913-0.932), and southeast direction, 0.87 mV (AUROC, 0.906; 95% CI, 0.895-0.917) in HD-16; and across 0.83 mV (AUROC, 0.917; 95% CI, 0.911-0.924), along 0.46 mV (AUROC, 0.890; 95% CI, 0.883-0.897), northeast direction 0.89 mV (AUROC, 0.923; 95% CI, 0.917-0.929), and southeast direction 0.83 mV (AUROC, 0.913; 95% CI, 0.906-0.920) in HD-32. Significant differences in AUROC were seen between HD-16 along versus across (P=0.002), HD-16 northeast direction versus southeast direction (P=0.01), HD-32 northeast direction versus southeast direction (P<0.0001), and HD-16 along versus HD-32 along (P=0.006). The AUROC was significantly larger (P<0.01) when only the best points on each given site were selected for analysis, compared with when all points were used. CONCLUSIONS: Spacing and orientation of bipoles impacts the accuracy of scar detection. Optimal threshold specific to each bipolar configuration should be determined. Selecting one best voltage point among multiple points projected on the same surface is also critical on the Ensite-system to increase the accuracy of scar-mapping.
Fichier principal
Vignette du fichier
Takigawa - CIRCEP.119.007158.pdf (490.75 Ko) Télécharger le fichier
Origine : Publication financée par une institution

Dates et versions

hal-03033103 , version 1 (26-10-2021)

Identifiants

Citer

Masateru Takigawa, Jatin Relan, Takeshi Kitamura, Claire A. Martin, Steven Kim, et al.. Impact of Spacing and Orientation on the Scar Threshold With a High-Density Grid Catheter. Circulation. Arrhythmia and electrophysiology, 2019, 12 (9), ⟨10.1161/CIRCEP.119.007158⟩. ⟨hal-03033103⟩

Collections

INSERM
26 Consultations
62 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More