Eurados 2016 intercomparison exercise of eye lens dosemeters - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Radiation Protection Dosimetry Année : 2018

Eurados 2016 intercomparison exercise of eye lens dosemeters

Résumé

In the context of a new annual eye lens dose limit for occupational exposure equal to 20mSv, European Radiation Dosimetry Group (EURADOS) organized an intercomparison dedicated to eye lens dosemeters, including photon and beta radiations. The objective was to complete the first intercomparison recently organized by EURADOS for photons and to update the overview of eye lens dosemeters available in Europe. The dosemeters provided by the 22 participants coming from 12 countries were all composed of thermoluminescent detectors. The dosemeters were irradiated with photon and beta fields defined in relevant standards. The results, provided by participants in terms of Hp(3), were compared to the reference delivered doses. Results are globally satisfactory for photons since 90% of the data are in accordance to the ISO 14146 standard requirements. The respective values for betas stress the fact that dosemeters designed for Hp(0.07) are not suitable to monitor the eye lens dose in case of betas. © The Author(s) 2018.
Fichier non déposé

Dates et versions

hal-02881792 , version 1 (26-06-2020)

Identifiants

Citer

I. Clairand, R. Behrens, M. Brodecki, E. Carinou, J. Domienik-Andrzejewska, et al.. Eurados 2016 intercomparison exercise of eye lens dosemeters. Radiation Protection Dosimetry, 2018, 182 (3), pp.317-322. ⟨10.1093/RPD/NCY067⟩. ⟨hal-02881792⟩

Collections

IRSN
15 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Mastodon Facebook X LinkedIn More