Calibration of thermocouple-based scanning thermal microscope in active mode (2ω method)
Résumé
ABSTRACT
We present a procedure dedicated to the calibration of a scanning thermal microscopy probe operated in an active mode and a modulated
regime especially for the measurement of solid material thermal conductivities. The probe used is a microthermocouple wire mounted on a
quartz tuning fork. Measurements on reference samples are performed successively in vacuum and ambient air conditions revealing a clear
difference in the dependence of the thermal interaction between the probe and the sample on the sample properties. Analytical modeling
based on the resolution of the heat equation in the wire probe and a description of the thermal interaction using a network of thermal
conductances are used to fit experimental data and analyze this difference. We have experimentally verified that the effective thermal contact
radius of the probe tip depends on the sample thermal conductivity in ambient conditions, whereas it remains constant in vacuum. We have
defined the measurement range of the technique based on the decrease in the probe sensitivity at high thermal conductivities. Considering
the experimental noise of our electrical device, it is shown that the maximum measurable value of thermal conductivity is near 23Wm−1 K−1
in vacuum and 37 W m−1 K−1 in ambient air conditions. Moreover, the lowest uncertainties are obtained for thermal conductivities below
5Wm−1 K−1 typically.