Calibrated force measurement in Atomic Force Microscopy using the Transient Fluctuation Theorem - Archive ouverte HAL Accéder directement au contenu
Pré-Publication, Document De Travail Année : 2020

Calibrated force measurement in Atomic Force Microscopy using the Transient Fluctuation Theorem

Résumé

The Transient Fluctuation Theorem is used to calibrate an Atomic Force Microscope by measuring the fluctuations of the work performed by a time dependent force applied between a colloïdal probe and the surface. From this measure one can easily extract the value of the interaction force and the relevant parameters of the cantilever. The results of this analysis are compared with those obtained by standard calibration methods. a) present adress: ISIS, Univ.
Fichier principal
Vignette du fichier
ArticleTF_V13.pdf (1.1 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-02561947 , version 1 (04-05-2020)
hal-02561947 , version 2 (15-12-2020)

Identifiants

Citer

Samuel Albert, Aubin Archambault, Artyom Petrosyan, Caroline Crauste-Thibierge, Ludovic Bellon, et al.. Calibrated force measurement in Atomic Force Microscopy using the Transient Fluctuation Theorem. 2020. ⟨hal-02561947v1⟩
118 Consultations
101 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More