Coupling between a calix[6]arene-based chromatography column and ICP-MS for on-line actinide analysis
Résumé
A method based on the coupling of a calixarene column used for the preconcentration of actinides to a quadrupole ICP-MS was developed for the separation of actinides and their on-line detection. First, a flow injection system was developed for aqueous U, Pu and Am on-line ICP-MS analysis. Then, chromatography parameters such as the column diameter and flow rates were optimised using the theoretical plate model. The detection limits achieved are 0.6 ng (7 μBq) for 238U, 16 pg (37 mBq) for 239Pu and 0.21 pg (27 mBq) for 241Am. The signal improvement achieved by this coupling method was discussed and compared to "dilute and shoot" analysis. Finally, this method was validated on certified reference materials of 238U, 239Pu and 241Am, at concentration levels close to the detection limits. © The Royal Society of Chemistry 2016.