A comprehensive theoretical comparison of proton imaging set-ups in terms of spatial resolution

Résumé

We present a comprehensive analytical comparison of four types of proton imaging set-ups and, to this end, develop a mathematical framework to calculate the width of the uncertainty envelope around the most likely proton path depending on set-up geometry, detector properties, and proton beam parameters. As a figure of merit for the spatial resolution achievable with each set-up, we use the frequency f10% at which the modular transfer function of a density step decreases below 10%. We verify the analytical results with Monte Carlo simulations. We find that set-ups which track the angle and position of individual protons in front of and behind the phantom would yield an average spatial resolution of 0.3–0.35 lp mm−1 assuming realistic geometric parameters (i.e. 30–40 cm distance between detector and phantom, 15–20 cm phantom thickness). For set-ups combining pencil beam scanning with either a position sensitive detector, e.g. an x-ray flat panel, or with a position insensitive detector, e.g. a range telescope, we find an average spatial resolution of about 0.1 lp mm−1 for an 8 mm FWHM beam spot size. The pixel information improves the spatial resolution by less than 10%. In both set-up types, performance can be significantly improved by reducing the pencil beam size down to 2 mm FWHM. In this case, the achievable spatial resolution reaches about 0.25 lp mm−1. Our results show that imaging setups combining double scattering with a pixel detector can provide sufficient spatial resolution only under very stringent conditions and are not ideally suited for computed tomography applications. We further propose a region-of-interest method for set-ups with a pixel detector to filter out protons which have undergone nuclear reactions and discuss the impact of tracker detector uncertainties on the most likely path.
Fichier principal
Vignette du fichier
Krah2018_authorsmanuscript.pdf (3.61 Mo) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-01909706 , version 1 (26-11-2019)

Identifiants

Citer

N. Krah, Feriel Khellaf, Jean Michel Létang, S. Rit, I. Rinaldi. A comprehensive theoretical comparison of proton imaging set-ups in terms of spatial resolution. Physics in Medicine and Biology, 2018, 63 (13), ⟨10.1088/1361-6560/aaca1f⟩. ⟨hal-01909706⟩
165 Consultations
261 Téléchargements

Altmetric

Partager

More