Titanium-titanium modular neck for primary THA. Result of a prospective series of 170 cemented THA with a minimum follow-up of 5 years - Archive ouverte HAL
Article Dans Une Revue Orthopaedics & Traumatology: Surgery & Research Année : 2015

Titanium-titanium modular neck for primary THA. Result of a prospective series of 170 cemented THA with a minimum follow-up of 5 years

Résumé

Background: Although they have been in use since the end of the 1980s, modular titanium neck components are associated with a risk of wear or fracture, and their safety has recently become a subject of debate and has never been evaluated in a consecutive series of patients. The goal of this study was to evaluate: revision-free survival of these implants after a minimum follow-up of 5 years; clinical and radiographic results; and the potential complications associated with the use of modular titanium neck components. Hypothesis: The use of titanium modular neck on cemented titanium THA is safe at a minimum follow-up of 5 years. Patients and methods: Between January 2006 and December 2008, we prospectively followed 170 patients (170 hips) who underwent primary anatomical THA with a modular cemented titanium stem design implant. The indications were unilateral THA for primary (n = 160) or secondary (n = 10) hip osteoarthritis (aseptic osteonecrosis of the femoral head or hip dysplasia). Mean age of patients was 75.4 +/- 5.8 years old (52-85), and mean BMI was 26.1 +/- 4.5 kg/m(2) (16.6-42.1). Patients were operated on by a modified Watson-Jones anterolateral approach based on preoperative 2D planning. All patients underwent annual clinical and radiological follow-up by an independent observer. Results: At a mean follow-up of 71 +/- 8 months (60-84), 5 patients died and 7 were lost to follow-up. There was no revision of THA after a maximum follow-up of 84 months. The Harris score improved significantly from 50.4 +/- 11.3 (0-76) preoperatively to 84.5 +/- 15.2 (14-100) at the final follow-up. There was no difference in postoperative femoral offset or the position of the center of rotation compared to the opposite side. On the other hand, the neck-shaft angle (NSA) and limb length were corrected (2 +/- 5 degrees [-11 to +14] and 2.16 +/- 3.6 mm [-7.4 to +12.7 mm]) respectively. Fifteen patients (9%) had limb length discrepancies of more than 5 mm and 4 patients (2%) of more than 10 mm. There were no complications due to the modular implant design. Discussion: Our study suggests that the use of cemented titanium implants with a modular titanium stem is safe at a follow-up of 5 years. The modular design does not prevent limb length discrepancies but restores femoral offset. Level of evidence: IV: prospective, non-comparative study. (C) 2015 Elsevier Masson SAS. All rights reserved.

Domaines

Chirurgie

Dates et versions

hal-01449321 , version 1 (30-01-2017)

Identifiants

Citer

M. Ollivier, S. Parratte, A. Galland, A. Lunebourg, X. Flecher, et al.. Titanium-titanium modular neck for primary THA. Result of a prospective series of 170 cemented THA with a minimum follow-up of 5 years. Orthopaedics & Traumatology: Surgery & Research, 2015, 101 (2), pp.137-142. ⟨10.1016/j.otsr.2014.12.013⟩. ⟨hal-01449321⟩
50 Consultations
0 Téléchargements

Altmetric

Partager

More