Assessment of Geant4 Prompt-Gamma Emission Yields in the Context of Proton Therapy Monitoring - Archive ouverte HAL
Article Dans Une Revue Frontiers in Oncology Année : 2016

Assessment of Geant4 Prompt-Gamma Emission Yields in the Context of Proton Therapy Monitoring

Résumé

Monte Carlo tools have been long used to assist the research and development of solutions for proton therapy monitoring. The present work focuses on the prompt-gamma emission yields by comparing experimental data with the outcomes of the current version of Geant4 using all applicable proton inelastic models. For the case in study and using the binary cascade model, it was found that Geant4 overestimates the prompt-gamma emission yields by 40.2 ± 0.3%, even though it predicts the prompt-gamma profile length of the experimental profile accurately. In addition, the default implementations of all proton inelastic models show an overestimation in the number of prompt gammas emitted. Finally, a set of built-in options and physically sound Geant4 source code changes have been tested in order to try to improve the discrepancy observed. A satisfactory agreement was found when using the QMD model with a wave packet width equal to 1.3 fm 2 .
Fichier principal
Vignette du fichier
fonc-06-00010.pdf (184.28 Ko) Télécharger le fichier
Origine Fichiers éditeurs autorisés sur une archive ouverte
Loading...

Dates et versions

hal-01272829 , version 1 (11-02-2016)

Identifiants

Citer

M. Pinto, Denis Dauvergne, Nicolas Freud, Jochen Krimmer, Jean Michel Létang, et al.. Assessment of Geant4 Prompt-Gamma Emission Yields in the Context of Proton Therapy Monitoring. Frontiers in Oncology, 2016, 6 (10), ⟨10.3389/fonc.2016.00010⟩. ⟨hal-01272829⟩
250 Consultations
272 Téléchargements

Altmetric

Partager

More