NiTi based stent for the treatment of acute urinary retention due to benign prostatic hyperplasia: a preliminary study on NiTi wires and tubes under pure bending - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Computer Methods in Biomechanics and Biomedical Engineering Année : 2014

NiTi based stent for the treatment of acute urinary retention due to benign prostatic hyperplasia: a preliminary study on NiTi wires and tubes under pure bending

Résumé

Benign prostatic hyperplasia (BPH) remains an incompletely understood disease, mostly striking old men. As an alternative to catheterization, an innovative NiTi shape memory alloy (SMA) based stent, made of a helical wire inserted into a helical tube was developed. In particular, this stent presents at least two distinct successive shape memories during heating, allowing an extension phase to release the urethra, and a contraction phase for an easy removal. This paper presents the stent mechanism, and the first mechanical characterization of the NiTi wires and tubes, under pure bending.
Fichier principal
Vignette du fichier
CMBBE SB 2014 Antherieu.pdf (430.53 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-01108621 , version 1 (23-01-2015)

Identifiants

Citer

Gabriel Antherieu, Nathanaël Connesson, Yohan Payan, Denis Favier, Pierre Mozer. NiTi based stent for the treatment of acute urinary retention due to benign prostatic hyperplasia: a preliminary study on NiTi wires and tubes under pure bending. Computer Methods in Biomechanics and Biomedical Engineering, 2014, 17, pp.190 - 191. ⟨10.1080/10255842.2014.931679⟩. ⟨hal-01108621⟩
1620 Consultations
128 Téléchargements

Altmetric

Partager