Registration-based geometric calibration of industrial X-ray tomography system

Résumé

This paper presents a general and efficient method for geometrical calibration of industrial X-ray tomography systems without the use of any phantom. In the field of nondestructive testing (NDT), an accurate numericalmodel of the object is usually available. By extending the classical volume-to-image registration framework, the method estimates both the parameters related to the position of the object, and those related to the positions and the orientations of the X-ray source, the detector and the rotation axis. Only using the projections data and the numericalmodel of the object, the values of the geometrical parameters are estimated with a good accuracy. Numerical experiments show that the method provides good results, even if the initialization of the process is far from the real values.
Fichier non déposé

Dates et versions

hal-00878581 , version 1 (30-10-2013)

Identifiants

Citer

Ismail Ben Tekaya, Valerie Kaftandjian, Fanny Buyens, Sylvie Sevestre-Ghalila, Samuel Legoupil. Registration-based geometric calibration of industrial X-ray tomography system. IEEE Transactions on Nuclear Science, 2013, 60, pp.1-8. ⟨10.1109/TNS.2013.2279675⟩. ⟨hal-00878581⟩