Ligament imbalance metrics and an autonomous measurement system for post TKA - Archive ouverte HAL
Communication Dans Un Congrès Année : 2010

Ligament imbalance metrics and an autonomous measurement system for post TKA

Résumé

This paper presents possible telemetry system architectures to evaluate the collateral ligaments imbalance for post Total Knee Arthroplasty (TKA) scenario. Two metrics are taken into account: the center of pressure and the net moments (varus-valgus and anterior posterior). Digital and analog system architectures are described for the COP metric and some are evaluated using high level models and experimental measurement. The architectures are designed considering that the implanted must be compliant with MICS standard and will be self-powered, implying a power source below 2 mW. In terms of hardware architecture, the paper shows that there is an interest for the analog architecture in terms of surface area and power consumption.

Domaines

Electronique
Fichier non déposé

Dates et versions

hal-00540867 , version 1 (29-11-2010)

Identifiants

  • HAL Id : hal-00540867 , version 1

Citer

Deepak Gupta, Cyril Lahuec, Matthieu Arzel, Chafiaa Hamitouche-Djabou, Michel Jezequel. Ligament imbalance metrics and an autonomous measurement system for post TKA. 32nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Aug 2010, Buenos Aires, Argentina. pp.6417 - 6420. ⟨hal-00540867⟩