A novel conductometric sensor based on a PVC membrane containing nonactin for ammonium determination - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue International Journal of Environmental Analytical Chemistry Année : 2009

A novel conductometric sensor based on a PVC membrane containing nonactin for ammonium determination

Résumé

This work presents a new sensor for ammonium determination based on interdigitated conductometric thin-film planar electrodes. It was fabricated by including nonactin in a plasticised poly(vinyl chloride) membrane deposited on the sensitive area of the transducer. The effects of pH, buffer concentration, the interferences of Na+, K+ and Ca2+ ions, the effect of ionic strength and of temperature were investigated. The linear working range and sensitivity of the biosensor were also determined. The sensitivity of the ammonium detection was around 6.2 S M-1 [image omitted] and the detection limit was 2 10-5 M. The biosensor exhibits the advantages of a simple fabrication procedure, good sensor-to-sensor reproducibility, wide dynamic range and good storage stability (12 weeks).
Fichier principal
Vignette du fichier
LY2009-PUB00026644.pdf (173.27 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-00508423 , version 1 (03-08-2010)

Identifiants

Citer

B. Khadro, P. Namour, F. Bessueille, D. Leonard, N. Jaffrezic Renault. A novel conductometric sensor based on a PVC membrane containing nonactin for ammonium determination. International Journal of Environmental Analytical Chemistry, 2009, 89 (1), p. 11 - p. 19. ⟨10.1080/03067310802339785⟩. ⟨hal-00508423⟩
166 Consultations
297 Téléchargements

Altmetric

Partager

Gmail Mastodon Facebook X LinkedIn More