Article Dans Une Revue Frontiers in Medicine Année : 2025

Specific reaction conditions for efficient automated $^{68}$Ga-radiolabeling of the FAP-2286 pseudopeptide on a GAIA® synthesizer

Résumé

Introduction: Automated radiolabeling of gallium-68-labeled experimental radiopharmaceuticals is crucial for ensuring high reproducibility and regulatory compliance in clinical settings. FAP-2286, a promising DOTApseudopeptide targeting the tumor microenvironment, has demonstrated superior tumor retention compared to quinoline-based analogs, making it an attractive theranostic agent. This study aimed to optimize and automate the preparation of [$^{68}$Ga]Ga-FAP-2286 on the GAIA synthesizer, ensuring high radiochemical purity (RCP) and radiochemical yield (RCY). Methods: Manual radiolabeling assays were initially performed to identify optimal reaction conditions, varying buffer, antioxidant, vector amount, heating time, and purification methods. The selected conditions were then adapted to an automated protocol using a GAIA module. A strong cation exchange (SCX) cartridge for $^{68}$Ga pre-concentration and a solid-phase extraction (SPE) step for final purification were included in the process. RCY, RCP, and stability over 4 h were assessed using radio-HPLC and radio-TLC. Additionally, the applicability of the optimized automated method was evaluated for 3BP-3940, a structurally related pseudopeptide. Results: Initial optimization studies identified sodium acetate buffer 0.1 M with methionine as an antioxidant, 25 µg of FAP-2286, and a 4-min heating time as the best manual radiolabeling conditions, achieving a RCP > 98%. In the automated synthesis, adjustments were made, including doubling the vector amount and extending heating to 9 min, resulting over three test-batches in a moderate RCY of 59.85 ± 3.73% and a RCP just over 94% up to 4 h after the end of synthesis. Importantly, the method was successfully transposed to [$^{68}$Ga]Ga-3BP-3940, yielding better RCY (75.62 ± 11.76%), RCP and stability profiles (>95.95% over 4 h). Conclusion: This study established a robust, automated protocol for the synthesis of [$^{68}$Ga]Ga-FAP-2286, ensuring high purity, reproducibility, and compatibility with clinical applications. The method's successful adaptation to 3BP-3940 highlights its versatility for such radiopharmaceuticals, supporting the broader implementation of automated theranostic agent production in nuclear medicine.

Fichier principal
Vignette du fichier
fmed-1-1628158.pdf (1.55 Mo) Télécharger le fichier
Origine Fichiers éditeurs autorisés sur une archive ouverte
Licence

Dates et versions

hal-05177369 , version 1 (22-07-2025)

Licence

Identifiants

Citer

Maissa Ammour, Jade Torchio, Stéphane C Renaud, Léa Rubira, Cyril Fersing. Specific reaction conditions for efficient automated $^{68}$Ga-radiolabeling of the FAP-2286 pseudopeptide on a GAIA® synthesizer. Frontiers in Medicine, 2025, 12, pp.1628158. ⟨10.3389/fmed.2025.1628158⟩. ⟨hal-05177369⟩
109 Consultations
195 Téléchargements

Altmetric

Partager

  • More