Real-time acousto-optic imaging using a high peak power long-pulsed illumination - Archive ouverte HAL
Article Dans Une Revue Optics Letters Année : 2024

Real-time acousto-optic imaging using a high peak power long-pulsed illumination

Résumé

Acousto-optic imaging (AOI) of absorbing objects embedded in highly scattering media remains challenging since the detectable signal suitable for image reconstruction is weak. To increase the detected signals to a level required by live biological applications, we designed a high peak power quasi-continuous laser source based on the coherent combination of two pulsed amplifiers, delivering 100 µs-long pulses with a 9 W peak power at a 100 Hz repetition rate while maintaining an average power below 100 mW; jointly used with a digital holographic detection which maximises the amount of collected signal, we demonstrate for the first time the optical imaging of 2-cm-thick highly scattering media (µ ′ s ∼ 10 cm -1 ) at near video frame rate (0.2 Hz) using Fourier Transform-AOI.
Fichier principal
Vignette du fichier
ViewAcceptedManuscript.pdf (18.31 Mo) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Licence

Dates et versions

hal-04718257 , version 1 (02-10-2024)

Licence

Identifiants

Citer

François Fligliolia, Qin Liu, Sylvie Janicot, Patrick Georges, Gaëlle Lucas-Leclin, et al.. Real-time acousto-optic imaging using a high peak power long-pulsed illumination. Optics Letters, In press, ⟨10.1364/OL.528953⟩. ⟨hal-04718257⟩
72 Consultations
26 Téléchargements

Altmetric