Cross-detector scatter issues in dual synchronous tomography: An affine projection correction protocol - Archive ouverte HAL
Article Dans Une Revue Tomography of Materials and Structures Année : 2024

Cross-detector scatter issues in dual synchronous tomography: An affine projection correction protocol

Résumé

Dual-beam x-ray tomography systems are paving the way for new experimental procedures, such as multi-resolution and multi-energy imaging, where synchronous acquisitions are essential. However, in such systems, cross-detector scatter between the detecting devices can occur as the two beamlines operate simultaneously. This paper proposes a new affine image transformation model of each projection to correct for these cross-detector scatter issues. A toy tomography test case is presented to assess the feasibility and performance of the proposed correction method.