In-situ measurements of NH 3 : instrument performance and applicability - Archive ouverte HAL
Article Dans Une Revue Atmospheric Measurement Techniques Discussions Année : 2022

In-situ measurements of NH 3 : instrument performance and applicability

Marsailidh Twigg
Augustinus Berkhout
Nicholas Cowan
  • Fonction : Auteur
Enrico Dammers
  • Fonction : Auteur
Volker Ebert
  • Fonction : Auteur
Marty Haaima
  • Fonction : Auteur
Christoph Häni
Lewis John
  • Fonction : Auteur
Matthew Jones
  • Fonction : Auteur
Bjorn Kamps
  • Fonction : Auteur
John Kentisbeer
  • Fonction : Auteur
Thomas Kupper
Sarah Leeson
  • Fonction : Auteur
Daiana Leuenberger
  • Fonction : Auteur
Nils Lüttschwager
  • Fonction : Auteur
Ulla Makkonen
  • Fonction : Auteur
Nicholas Martin
  • Fonction : Auteur
David Missler
  • Fonction : Auteur
Duncan Mounsor
  • Fonction : Auteur
Albrecht Neftel
  • Fonction : Auteur
Chad Nelson
  • Fonction : Auteur
Eiko Nemitz
Rutger Oudwater
  • Fonction : Auteur
Celine Pascale
Jean-Eudes Petit
Andrea Pogany
  • Fonction : Auteur
Jörg Sintermann
  • Fonction : Auteur
Amy Stephens
  • Fonction : Auteur
Mark Sutton
Yuk Tang
Rens Zijlmans
  • Fonction : Auteur
Christine Braban
Bernhard Niederhauser
  • Fonction : Auteur

Résumé

Abstract. Ammonia (NH3) in the atmosphere affects both the environment and human health. It is therefore increasingly recognised by policy makers as an important air pollutant that needs to be mitigated. In order to understand the effectiveness of abatement strategies, routine NH3 monitoring is required. Current reference protocols, developed in the 1990s, use daily samplers with offline analysis but there have been a number of technologies developed since, which may be applicable for high time resolution routine monitoring of NH3 at ambient concentrations. The following study is a comprehensive field intercomparison held over an intensively managed grassland in South East Scotland using currently available methods that are reported to be suitable for routine monitoring of ambient NH3. In total 13 instruments took part in the field study. The instruments include: an online ion chromatography system (MARGA, Metrohm-Applikon,NL), two wet chemistry continuous flow analysis systems (AiRRmonia, Mechatronics, NL), a photoacoustic spectrometer (NH3 monitor, LSE, NL), two mini Differential Optical Absorption Spectrometers (miniDOAS; NTB Interstate University of Applied Sciences Buchs, now part of "Eastern Switzerland University of Applied Sciences, CH and RIVM, NL), as well as seven spectrometers using cavity enhanced techniques: a Quantum Cascade Laser Absorption Spectrometer (QCLAS, Aerodyne, Inc. US), Picarro G2103 Analyzer (Picarro US), Economical NH3 Analyser (Los Gatos Research, US), Tiger-i 2000 (Tiger Optics, US) and LaserCEM® gas analyser (AP2E, FR). Assessments of the instruments’ precision at low concentrations (< 10 ppb) and at elevated concentrations (maximum reported concentration of 282 ppb) were undertaken. At elevated concentrations all instruments performed well on precision (r2 > 0.75). At concentrations below 10 ppb however, instruments fell into two distinct groups and the duplicate instruments, miniDOAS, AiRRmonia, LGR and Picarro were split across the two groups. It was found that identical instruments performed differently at low concentrations, highlighting the impact of the setup, inlet design and operation of the instrument used. Accuracy in determining absolute concentrations in the field was assessed using a calibration-free CRDS Optical Gas Standard (OGS, PTB, DE), serving as an instrumental reference standard. Accuracy was also assessed using well established metrological standards for calibration gases, i) a permeation system (ReGaS1, METAS, CH) and ii) Primary Standard gas Mixtures (PSMs) prepared by gravimetry (NPL, UK). This study showed that though the OGS good performance with respect to sensitivity and linearity with reference gas standards, this in itself is not enough for the OGS to be a field reference standard because a closed path spectrometer has limitations due to losses to surfaces in sampling NH3, which need to be taken into account. Overall, the instruments studied performed well against the standard gases but we note that not every instrument could be calibrated using gas standards due to incompatible inlet designs and limitations in the gas flow rates of the standards. This work provides evidence that though NH3 instrumentation have greatly progressed in measurement precision, there is still further work required to quantify the accuracy of these systems under field conditions. It is the recommendation of this study that the use of instruments for routine monitoring of NH3 needs to be set out in standard operating protocols for inlet set-up, calibration and routine maintenance, in order for datasets to be comparable.

Dates et versions

hal-03968179 , version 1 (01-02-2023)

Identifiants

Citer

Marsailidh Twigg, Augustinus Berkhout, Nicholas Cowan, Sabine Crunaire, Enrico Dammers, et al.. In-situ measurements of NH 3 : instrument performance and applicability. Atmospheric Measurement Techniques Discussions, 2022, ⟨10.5194/amt-2022-107⟩. ⟨hal-03968179⟩
41 Consultations
0 Téléchargements

Altmetric

Partager

More