Repeatability and reproducibility of a handheld quantitative G6PD diagnostic. - Archive ouverte HAL
Article Dans Une Revue PLoS Neglected Tropical Diseases Année : 2022

Repeatability and reproducibility of a handheld quantitative G6PD diagnostic.

Greg Bizilj
  • Fonction : Auteur
Gonzalo J Domingo
  • Fonction : Auteur
Sampa Pal
  • Fonction : Auteur

Résumé

Background The introduction of novel short course treatment regimens for the radical cure of Plasmodium vivax requires reliable point-of-care diagnosis that can identify glucose-6-phosphate dehydrogenase (G6PD) deficient individuals. While deficient males can be identified using a qualitative diagnostic test, the genetic make-up of females requires a quantitative measurement. SD Biosensor (Republic of Korea) has developed a handheld quantitative G6PD diagnostic (STANDARD G6PD test), that has approximately 90% accuracy in field studies for identifying individuals with intermediate or severe deficiency. The device can only be considered for routine care if precision of the assay is high. Methods and findings Commercial lyophilised controls (ACS Analytics, USA) with high, intermediate, and low G6PD activities were assessed 20 times on 10 Biosensor devices and compared to spectrophotometry (Pointe Scientific, USA). Each device was then dispatched to one of 10 different laboratories with a standard set of the controls. Each control was tested 40 times at each laboratory by a single user and compared to spectrophotometry results. When tested at one site, the mean coefficient of variation (CV) was 0.111, 0.172 and 0.260 for high, intermediate, and low controls across all devices respectively; combined G6PD Biosensor readings correlated well with spectrophotometry (rs = 0.859, p<0.001). When tested in different laboratories, correlation was lower (rs = 0.604, p<0.001) and G6PD activity determined by Biosensor for the low and intermediate controls overlapped. The use of lyophilised human blood samples rather than fresh blood may have affected these findings. Biosensor G6PD readings between sites did not differ significantly (p = 0.436), whereas spectrophotometry readings differed markedly between sites (p<0.001). Conclusions Repeatability and inter-laboratory reproducibility of the Biosensor were good; though the device did not reliably discriminate between intermediate and low G6PD activities of the lyophilized specimens. Clinical studies are now required to assess the devices performance in practice.
Fichier principal
Vignette du fichier
islandora_156519.pdf (1.51 Mo) Télécharger le fichier
Origine Fichiers éditeurs autorisés sur une archive ouverte

Dates et versions

hal-03790033 , version 1 (27-09-2022)

Licence

Identifiants

Citer

Benedikt Ley, Ari Winasti Satyagraha, Mohammad Golam Kibria, Jillian Armstrong, Germana Bancone, et al.. Repeatability and reproducibility of a handheld quantitative G6PD diagnostic.. PLoS Neglected Tropical Diseases, 2022, 16 (2), pp.e0010174. ⟨10.1371/journal.pntd.0010174⟩. ⟨hal-03790033⟩
166 Consultations
80 Téléchargements

Altmetric

Partager

More