How to assess depth-of-field extension strategies applied to single-molecule super-resolution microscopes?
Résumé
Phase masks used to extend the depth-of-field (DoF) of localization microscopes are often designed to increase localization accuracy despite a loss in detection probability. We propose a method to optimize DoF extended single-molecule localization techniques taking into account the trade-off between detection probability and localization accuracy.
Origine | Fichiers produits par l'(les) auteur(s) |
---|