Comparison of Holmium:YAG and Thulium Fiber Lasers on the Risk of Laser Fiber Fracture
Résumé
Objectives: To compare the risk of laser fiber fracture between Ho:YAG laser and Thulium Fiber Laser (TFL) with different laser fiber diameters, laser settings, and fiber bending radii.
Methods: Lengths of 200, 272, and 365 μm single use fibers were used with a 30 W Ho:YAG laser and a 50 W Super Pulsed TFL. Laser fibers of 150 µm length were also tested with the TFL only. Five different increasingly smaller bend radii were tested: 1, 0.9, 0.75, 0.6, and 0.45 cm. A total of 13 different laser settings were tested for the Ho:YAG laser: six fragmentation settings with a short pulse duration, and seven dusting settings with a long pulse duration. A total of 33 different laser settings were tested for the TFL. Three laser settings were common two both lasers: 0.5 J × 12 Hz, 0.8 J × 8 Hz, 2 J × 3 Hz. The laser was activated for 5 min or until fiber fracture. Each measurement was performed ten times.
Results: While fiber failures occurred with all fiber diameters with Ho:YAG laser, none were reported with TFL. Identified risk factors of fiber fracture with the Ho:YAG laser were short pulse and high energy for the 365 µm fibers (p = 0.041), but not for the 200 and 272 µm fibers (p = 1 and p = 0.43, respectively). High frequency was not a risk factor of fiber fracture. Fiber diameter also seemed to be a risk factor of fracture. The 200 µm fibers broke more frequently than the 272 and 365 µm ones (p = 0.039). There was a trend for a higher number of fractures with the 365 µm fibers compared to the 272 µm ones, these occurring at a larger bend radius, but this difference was not significant.
Conclusion: TFL appears to be a safer laser regarding the risk of fiber fracture than Ho:YAG when used with fibers in a deflected position.
Origine | Publication financée par une institution |
---|