Accurate and High-Precision Measurement of Lithium Isotopes in Two Reference Materials by MC-ICP-MS

Résumé

We report here a newly developed method for measurement of Li isotopes by use of multi-collector ICP-MS (Neptune) allowing rapid and high precision determination of Li isotope ratios at low levels of lithium (15–20 ng). The lithium reference sample solution IRMM-016 was analysed over a period of ten months with an external reproducibility of 0.24% (2s, n = 52). Chemical separation of Li from matrix was performed on the seawater sample IRMM BCR-403, for which a mean δ7Li value of + 31.0 ± 0.1 % (2s/√n, n = 31) was obtained. This mean value is in good agreement with those previously published for other seawater samples. BCR-403 seawater being readily available, we propose that this seawater sample be used as a reference sample for Li isotope measurements.

Dates et versions

hal-03154184 , version 1 (27-02-2021)

Identifiants

Citer

Romain Millot, Catherine Guerrot, Nathalie Vigier. Accurate and High-Precision Measurement of Lithium Isotopes in Two Reference Materials by MC-ICP-MS. Geostandards and Geoanalytical Research, 2004, 28 (1), pp.153-159. ⟨10.1111/j.1751-908X.2004.tb01052.x⟩. ⟨hal-03154184⟩