Simultaneous proton density, T1 , T2 , and flip‐angle mapping of the brain at 7 T using multiparametric 3D SSFP imaging and parallel‐transmission universal pulses

Résumé

Purpose: Performing simultaneous quantitative MRI at ultra-high field is challenging, as B0 and B1 + heterogeneities as well as Specific Absorption Rate increase. Too large deviations of flip angle from the target can induce biases and impair signal-to-noise ratio in the quantification process. In this work, we use calibration-free parallel transmission, a dedicated pulse sequence parameter optimization and signal fitting to recover 3D proton density, flip angle, T1 and T2 maps over the whole brain, in a clinically suitable time. Methods: Eleven optimized contrasts were acquired with an unbalanced Steady-State Free Precession sequence by varying flip angle amplitude and radiofrequency phase cycling increment, at a 1.0x1.0x3.0mm 3 resolution. Acquisition time was of 16min36sec for the whole brain. Parallel transmission and Universal Pulses were used to mitigate B1 + heterogeneity to improve the results' reliability over six healthy volunteers (3 females/males, age 22.6±2.7 years-old). Quantification of the physical parameters was performed by fitting acquired contrasts to the simulated ones using the Bloch-Torrey equations with a realistic diffusion coefficient. Results: Whole-brain 3D maps of effective flip angle, PD and relaxation times were estimated. Parallel transmission improved the robustness of the results at 7T. Results were in accordance with literature and with measurements from standard methods. Conclusion: These preliminary results show robust PD, FA, T1 and T2 map retrieval. Other parameters, such as Apparent Diffusion Coefficient, could be assessed. With further optimization in the acquisition, scan time could be reduced and spatial resolution increased to bring this multi-parametric quantification method to clinical research routine at 7-tesla.
Fichier principal
Vignette du fichier
LL200528_Quics_accepted_manuscript_hal.pdf (3.08 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-02897852 , version 1 (20-11-2020)

Identifiants

Citer

Lisa Leroi, Vincent Gras, Nicolas Boulant, Mathilde Ripart, Emilie Poirion, et al.. Simultaneous proton density, T1 , T2 , and flip‐angle mapping of the brain at 7 T using multiparametric 3D SSFP imaging and parallel‐transmission universal pulses. Magnetic Resonance in Medicine, 2020, 84 (6), pp.3286-3299. ⟨10.1002/mrm.28391⟩. ⟨hal-02897852⟩
123 Consultations
199 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More