Can the SCD test and terminal uridine nick-end labeling by flow cytometry technique (TUNEL/FCM) be used interchangeably to measure sperm DNA damage in routine laboratory practice?

Résumé

Background: Numerous tests have been proposed to evaluate sperm DNA integrity. To assess the sperm chromatin dispersion (SCD) test in an andrology laboratory, twenty-five men attending Clermont-Ferrand (France) University Hospital's Center for Reproductive Medicine were recruited. Sperm DNA damage was measured in the same semen samples using the SCD test and the Terminal Uridine Nick-end Labeling by flow cytometry technique (TUNEL/FCM) after density gradient centrifugation. Results: SCD test reliability between readings, readers or slides was clearly established with very high agreement between measurements (Intraclass correlation coefficient (ICC) at 0.97, 0.95 and 0.98 respectively). Despite very good agreement between the SCD test and TUNEL/FCM (ICC at 0.94), the SCD test tended to slightly but significantly underestimate DNA damage compared with TUNEL (p = 0.0127). This systematic difference between the two techniques was - 3.39 +/- 1.45% (mean +/- SE). Conclusions: Andrology laboratories using the SCD test to measure sperm DNA damage need to know that it appears to give slightly underestimated measurements compared to TUNEL/FCM. However, this systematic underestimation is very small in amplitude. Both techniques give almost perfectly congruent results. Our study underlines the importance for each laboratory to validate its method to assess sperm DNA damage before implementing it in routine andrology lab practice.
Fichier principal
Vignette du fichier
2020_Greze_Basic_Clin_Androl.pdf (836.53 Ko) Télécharger le fichier
Origine Publication financée par une institution

Dates et versions

hal-02493530 , version 1 (31-03-2021)

Licence

Identifiants

Citer

C. Greze, A. Guttmann, H. Pons-Rejraji, Marie-Paule Vasson, J. Lornage, et al.. Can the SCD test and terminal uridine nick-end labeling by flow cytometry technique (TUNEL/FCM) be used interchangeably to measure sperm DNA damage in routine laboratory practice?. Basic and clinical andrology, 2019, 29 (1), ⟨10.1186/s12610-019-0098-2⟩. ⟨hal-02493530⟩
79 Consultations
108 Téléchargements

Altmetric

Partager

More