Sensitive High Frequency AC Susceptometry in Magnetic Nanoparticle Applications - Archive ouverte HAL
Communication Dans Un Congrès Année : 2010

Sensitive High Frequency AC Susceptometry in Magnetic Nanoparticle Applications

Résumé

We report on the development of a sensitive high frequency susceptometer capable of measuring in the frequency range from 25 kHz up to 10 MHz with a volume susceptibility sensitivity of 3.5×10−5 at 100 kHz corresponding to about 0.3% of the measured AC susceptibility. In combination with the previous reported DynoMag system capable of measuring dynamic magnetic properties in the range from 1 Hz to 200 kHz we are thus able to measure dynamic magnetic properties between 1 Hz to 10 MHz with high magnetic sensitivity. We will show AC susceptometry applications and results within the fields of magnetic hyperthermia and dynamic magnetic characterization of magnetic nanoparticle system with different particle sizes and magnetic properties.
Fichier principal
Vignette du fichier
Sensitive High Frequency AC Susceptometry in Magnetic Nanoparticle Applications-final.pdf (414.22 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-02320524 , version 1 (18-10-2019)

Identifiants

Citer

Fredrik Ahrentorp, Andrea Astalan, Christian Jonasson, Jakob Blomgren, Qi Bin, et al.. Sensitive High Frequency AC Susceptometry in Magnetic Nanoparticle Applications. 8th International Conference on the Scientific and Clinical Applications of Magnetic Carriers, May 2010, Rostock, Germany. pp.213-223, ⟨10.1063/1.3530015⟩. ⟨hal-02320524⟩
109 Consultations
470 Téléchargements

Altmetric

Partager

More