A Low-Noise and Quasi-Ideal DC Current Source Dedicated to Four-Probe Low-Frequency Noise Measurements - Archive ouverte HAL
Article Dans Une Revue IEEE Transactions on Instrumentation and Measurement Année : 2020

A Low-Noise and Quasi-Ideal DC Current Source Dedicated to Four-Probe Low-Frequency Noise Measurements

Résumé

In this paper, a DC current source dedicated to four-probe low frequency noise measurements is presented. An output impedance value of 3 M with a maximum output current of 1.5 mA was achieved. The white noise level of the current source was measured to be 410-23 A 2 •Hz-1 with a corner frequency of 30 Hz. Thanks to its quasi-ideal behavior in terms of noise level and high output impedance value, this current source can be used to measure the intrinsic noise of materials using four-point measurements without any experimental or analytical corrections.

Domaines

Electronique
Fichier principal
Vignette du fichier
Combine Result.pdf (1.96 Mo) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-02273188 , version 1 (04-01-2021)

Identifiants

Citer

Jean-Marc Routoure, Sheng Wu, Carlo Barone, Laurence Méchin, Bruno Guillet. A Low-Noise and Quasi-Ideal DC Current Source Dedicated to Four-Probe Low-Frequency Noise Measurements. IEEE Transactions on Instrumentation and Measurement, 2020, 69 (1), pp.194. ⟨10.1109/TIM.2019.2890891⟩. ⟨hal-02273188⟩