Systematic Analysis of the Improvements in Magnetic Resonance Microscopy with Ferroelectric Composite Ceramics - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Advanced Materials Année : 2019

Systematic Analysis of the Improvements in Magnetic Resonance Microscopy with Ferroelectric Composite Ceramics

Résumé

The spatial resolution and signal-to-noise ratio (SNR) attainable in magnetic resonance microscopy (MRM) are limited by intrinsic probe losses and probe–sample interactions. In this work, the possibility to exceed the SNR of a standard solenoid coil by more than a factor-of-two is demonstrated theoretically and experimentally. This improvement is achieved by exciting the first transverse electric mode of a low-loss ceramic resonator instead of using the quasi-static field of the metal-wire solenoid coil. Based on theoretical considerations, a new probe for microscopy at 17 T is developed as a dielectric ring resonator made of ferroelectric/dielectric low-loss composite ceramics precisely tunable via temperature control. Besides the twofold increase in SNR, compared with the solenoid probe, the proposed ceramic probe does not cause static-field inhomogeneity and related image distortion.
Fichier principal
Vignette du fichier
adma.201900912-2.pdf (1.2 Mo) Télécharger le fichier
Origine : Publication financée par une institution
Loading...

Dates et versions

hal-02135592 , version 1 (17-04-2020)

Identifiants

Citer

Marine Moussu, Luisa Ciobanu, Sergej Kurdjumov, Elizaveta Nenasheva, Boucif Djemai, et al.. Systematic Analysis of the Improvements in Magnetic Resonance Microscopy with Ferroelectric Composite Ceramics. Advanced Materials, 2019, 31, pp.1900912. ⟨10.1002/adma.201900912⟩. ⟨hal-02135592⟩
166 Consultations
66 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More