BEEEP: B1-robust Energy Efficient Excitation Pulses - Archive ouverte HAL Accéder directement au contenu
Poster De Conférence Année : 2019

BEEEP: B1-robust Energy Efficient Excitation Pulses

Résumé

Synopsis This study introduces a new family of broadband B1-robust excitation (90°) pulses for MRI with large enough bandwidth (+/- 1 kHz) to account for static field inhomogeneities, and minimal energy deposition. RF pulses are designed with a regularized optimal control algorithm, which is able to adapt the pulse B1-robustness range to fit the coil limits in terms of peak amplitude and energy. In vitro acquisitions using an endoluminal-shaped RF transmit coil show comparable excitation profiles than BIR4 pulses, although BEEEP pulses deposit 5.2 times less energy. https://index.mirasmart.com/ISMRM2019/PDFfiles/4623.html

Domaines

Imagerie
Fichier non déposé

Dates et versions

hal-02130017 , version 1 (15-05-2019)

Identifiants

  • HAL Id : hal-02130017 , version 1

Citer

Eric van Reeth van Reeth, Hélène Ratiney, O. Beuf, Soukaina Kanice, Steffen Glaser, et al.. BEEEP: B1-robust Energy Efficient Excitation Pulses. 27th Annual meeting of the ISMRM, May 2019, Montreal, Canada. ⟨hal-02130017⟩