Mesoporous Titania-Coated Love Wave Sensors and FEM Model Applied to Viscosity Micro-Measurements - Archive ouverte HAL Accéder directement au contenu
Communication Dans Un Congrès Année : 2017

Mesoporous Titania-Coated Love Wave Sensors and FEM Model Applied to Viscosity Micro-Measurements

Résumé

This paper presents a Love wave sensor coated with a TiO2 mesoporous film. The main novelty is related to the modelling of the device and simulation by using FEM, which is a good way to take into account the physical properties of porous 3D-layers. Experimental devices have shown a good agreement with simulated results. Furthermore, experimental results with several Newtonian liquids are investigated. The sensitivity with a 300 nm thick porous layer was 30 times that of the bare device, with interesting dynamical issues to be further studied, giving rise to the great potentialities of such architectures for microrheological measurements.
Fichier non déposé

Dates et versions

hal-01719068 , version 1 (27-02-2018)

Identifiants

Citer

Ollivier Tamarin, Wassim Ouelhazi, Jean-Luc Lachaud, Hamida Hallil, Vincent Raimbault, et al.. Mesoporous Titania-Coated Love Wave Sensors and FEM Model Applied to Viscosity Micro-Measurements. IEEE Sensors, Oct 2017, Glasgow, United Kingdom. ⟨10.1109/ICSENS.2017.8234271⟩. ⟨hal-01719068⟩
280 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More