Artifact Correction for DVC Measurements Using a Laboratory X‐Ray Source - Archive ouverte HAL
Communication Dans Un Congrès Année : 2013

Artifact Correction for DVC Measurements Using a Laboratory X‐Ray Source

Résumé

The effects of temperature elevation of the X-ray tube of a lab tomograph during image acquisition are analyzed in terms of their induced biases for Digital Volume Correlation (DVC) from a series of reconstructed volumes acquired successively. Spurious dilatational strains are induced by temperature variations in the tomograph. If not accounted for, any quantitative kinematic measurement is impossible for strain levels below 0.5%. When corrected, it is possible to determine the elastic parameters of nodular graphite cast iron.
Fichier principal
Vignette du fichier
Adrien2013.pdf (127 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-01668877 , version 1 (26-07-2021)

Licence

Identifiants

Citer

Jérôme Adrien, Jean-Yves Buffiere, François Hild, Nathalie Limodin, Julien Réthoré, et al.. Artifact Correction for DVC Measurements Using a Laboratory X‐Ray Source. 2012 Annual Conference on Experimental and Applied Mechanics, 2012, Costa Mesa, United States. pp.217-220, ⟨10.1007/978-1-4614-4235-6_30⟩. ⟨hal-01668877⟩
118 Consultations
78 Téléchargements

Altmetric

Partager

More