Pulse-Shape Analysis of Ionization Signals in Cryogenic Ge Detectors for Dark Matter - Archive ouverte HAL
Communication Dans Un Congrès Année : 2016

Pulse-Shape Analysis of Ionization Signals in Cryogenic Ge Detectors for Dark Matter

Résumé

The detectors of the direct dark matter search experiment EDELWEISS consist of high-purity germanium crystals operated at cryogenic temperatures ( $\mathrm {{<}20\,mK}$ ) and low electric fields ( $\mathrm {{<}1\,V/cm}$ ). The surface discrimination is based on the simultaneous measurement of the charge amplitudes on different sets of electrodes. As the rise time of a charge signal strongly depends on the location of an interaction in the crystal, a time-resolved measurement can also be used to identify surface interactions. This contribution presents the results of a study of the discrimination power of the rise time parameter from a hot carrier transport simulation in combination with time-resolved measurements using an EDELWEISS-type detector in a test cryostat at ground level. We show the setup for the time-resolved ionization signal read-out in the EDELWEISS-III experiment and first results from data taking in the underground laboratory of Modane.
Fichier non déposé

Dates et versions

hal-01584671 , version 1 (09-09-2017)

Identifiants

Citer

N. Foerster, A. Broniatowski, K. Eitel, S. Marnieros, B. Paul, et al.. Pulse-Shape Analysis of Ionization Signals in Cryogenic Ge Detectors for Dark Matter. 16th International Workshop on Low Temperature Detectors, Jul 2015, Grenoble, France. pp.845-851, ⟨10.1007/s10909-016-1532-x⟩. ⟨hal-01584671⟩