Dual Frequency Band Annular Probe for Volumetric Pulse-echo Optoacoustic Imaging - Archive ouverte HAL Accéder directement au contenu
Communication Dans Un Congrès Année : 2015

Dual Frequency Band Annular Probe for Volumetric Pulse-echo Optoacoustic Imaging

François Varray
Didier Vray

Résumé

Optoacoustic (OA) pulse echo (PE) imaging is a hybridized modality that is capable of providing physiological information on the basis of anatomical structure. In this work, we propose a dual frequency band annular probe for backward mode volumetric PE/OA imaging. The performance of this design is evaluated based on the spatio-temporal impulse response, three dimensional steerability of the transducer and point spread function. Optimum settings for number of elements in each ring and maximum steering are suggested. The transducer design and synthetic array beamforming simulation are presented. The resolution performance and reconstruction capabilities are shown with the in-silico measurements.
Fichier principal
Vignette du fichier
Azizian 2015, Dual Frequency Band Annular Probe for Volumetric Pulse-Echo Optoacoustic Imaging-1.pdf (654.64 Ko) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte
Loading...

Dates et versions

hal-01279605 , version 1 (26-02-2016)

Identifiants

Citer

Mohammad Azizian Kalkhoran, François Varray, Didier Vray. Dual Frequency Band Annular Probe for Volumetric Pulse-echo Optoacoustic Imaging. 2015 ICU International Congress on Ultrasonics, May 2015, Metz, France. ⟨10.1016/j.phpro.2015.08.236⟩. ⟨hal-01279605⟩
110 Consultations
134 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More