Preliminary testing of GaN-based dosimeters for electron beam radiotherapy - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Radiation Protection Dosimetry Année : 2015

Preliminary testing of GaN-based dosimeters for electron beam radiotherapy

Résumé

The response of an implantable in vivo dosimetric system based on gallium nitride radioluminescence was investigated for electron beam radiotherapy using ELEKTA SLi and VARIAN Clinac 2100 CD Linear Accelerators. A bi-channel method has been implemented for fibre background rejection. The percentage depth dose (PDD) profiles were measured in polymethyl methacrylate for 6, 12 and 18 MeV electron beams. The PDD results were in excellent agreement with those measured with reference to ionisation chambers.
Fichier non déposé

Dates et versions

hal-01276553 , version 1 (19-02-2016)

Identifiants

Citer

A. Ismail, R. Wang, A. Chaikh, P. Pittet, J. Balosso. Preliminary testing of GaN-based dosimeters for electron beam radiotherapy. Radiation Protection Dosimetry, 2015, 165 (1-4), pp.488-491. ⟨10.1093/rpd/ncv056⟩. ⟨hal-01276553⟩