A Highly Sensitive Glucose Electrode Using Glucose Oxidase Collagen Film - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Bioelectrochemistry and Bioenergetics Année : 1978

A Highly Sensitive Glucose Electrode Using Glucose Oxidase Collagen Film

Résumé

Industrially reconstituted collagen films have shown excellent prop¬ erties for (3-D-glucose oxidase couplirg. Associated with a platinum anode for amperometric detection of hydrogen peroxide, these ei zymatic films form a very simple and easy to handle glucose electrode ; this device presents a very high sersitivity (ca. io~8 M) givirg resporses proportional to glucose concentration over 5 orders of magnitude.
Fichier principal
Vignette du fichier
ENPC00_AR_LEESU_0010.pdf (16.37 Mo) Télécharger le fichier
Origine : Accord explicite pour ce dépôt
Loading...

Dates et versions

hal-01179313 , version 1 (23-07-2015)

Identifiants

  • HAL Id : hal-01179313 , version 1

Citer

Daniel R. Thevenot, Pierre R. Coulet, Robert Sternberg, Daniele C. Gautheron. A Highly Sensitive Glucose Electrode Using Glucose Oxidase Collagen Film. Bioelectrochemistry and Bioenergetics, 1978, 5, pp.548-553. ⟨hal-01179313⟩