REAL -TIME WAVELET-BASED ALGORITHM FOR CARDIAC AND RESPIRATORY MRI GATING
Résumé
A real time algorithm for cardiac and respiratory gating, which only requires an ECG sensor, is proposed here. Three ECG electrodes are placed in such a manner that the modulation of the recorded ECG by the respiratory signal would be maximal; hence, given only one signal we can achieve both cardiac and respiratory MRI gating. First, an off-line learning phase based on wavelet decomposition is run to compute an optimal QRS filter. Afterwards, on one hand the QRS filter is used to accomplish R peak detection, and on the other, a low pass filtering process allows the retrieval of the respiration cycle so that the image acquisition sequences would be triggered by the R peaks only during the expiration phase.
Origine | Fichiers produits par l'(les) auteur(s) |
---|