An interferometric scanning microwave microscope and calibration method for sub-fF microwave measurements - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Review of Scientific Instruments Année : 2013

An interferometric scanning microwave microscope and calibration method for sub-fF microwave measurements

Résumé

We report on an adjustable interferometric set-up for Scanning Microwave Microscopy. This interferometer is designed in order to combine simplicity, a relatively flexible choice of the frequency of interference used for measurements as well as the choice of impedances range where the interference occurs. A vectorial calibration method based on a modified 1-port error model is also proposed. Calibrated measurements of capacitors have been obtained around the test frequency of 3.5 GHz down to about 0.1 fF. Comparison with standard vector network analyzer measurements is shown to assess the performance of the proposed system.
Fichier principal
Vignette du fichier
Dargent_2013_1.4848995.pdf (1.12 Mo) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte

Dates et versions

hal-00922585 , version 1 (02-06-2022)

Identifiants

Citer

Thomas Dargent, Kamel Haddadi, T. Lasri, N. Clément, D. Ducatteau, et al.. An interferometric scanning microwave microscope and calibration method for sub-fF microwave measurements. Review of Scientific Instruments, 2013, 84 (12), pp.123705. ⟨10.1063/1.4848995⟩. ⟨hal-00922585⟩
90 Consultations
52 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More