Finite Element Analysis of the Mechanical Performances of 8 Marketed Aortic Stent-Grafts
Résumé
Purpose: To assess numerically the flexibility and mechanical stresses undergone by stents and fabric of currently manufactured stent-grafts. Methods: Eight marketed stent-graft limbs (Aorfix, Anaconda, Endurant, Excluder, Talent, Zenith Flex, Zenith LP, and Zenith Spiral-Z) were modeled using finite element analysis. A numerical benchmark combining bending up to 180° and pressurization at 150 mmHg of the stent-grafts was performed. Stent-graft flexibility, assessed by the calculation of the luminal reduction rate, maximal stresses in stents, and maximal strains in fabric were assessed. Results: The luminal reduction rate at 90° was ‹<20% except for the Talent stent-graft. The rate at 180° was higher for Z-stented models (Talent, Endurant, Zenith, and Zenith LP; range 39%-78%) than spiral (Aorfix, Excluder, and Zenith Spiral-Z) or circular-stented (Anaconda) devices (range 14%-26%). At 180°, maximal stress was higher for Z-stented stent-grafts (range 370-622 MPa) than spiral or circular-stented endografts (range 177-368 MPa). At 90° and 180°, strains in fabric were low and did not differ significantly among the polyester stent-grafts (range 0.5%-7%), while the expanded polytetrafluoroethylene fabric of the Excluder stent-graft underwent higher strains (range 11%-18%). Conclusion: Stent design strongly influences mechanical performances of aortic stentgrafts. Spiral and circular stents provide greater flexibility, as well as lower stress values than Z-stents, and thus better durability.
Origine : Fichiers produits par l'(les) auteur(s)