Study of rough surface to decrease reverberation noise in ultrasonic imaging - Archive ouverte HAL Accéder directement au contenu
Communication Dans Un Congrès Année : 2012

Study of rough surface to decrease reverberation noise in ultrasonic imaging

Résumé

Rough back surface is investigated to decrease the reverberation noise in ultrasonic imaging. Silicon crystal is selected as the backing substrate of the ultrasonic transducer because rough structure is convenient to be fabricated on silicon substrate using microfabrication technologies. Different dimensions of rough boundaries are designed and simulated to scatter the undesired waves based on finite element method modeling. Transient analysis indicates that a rough surface whose dimension (including depth and width) is around 1.0 λ should be considered to scatter a majority of incident waves.
Fichier non déposé

Dates et versions

hal-00803090 , version 1 (21-03-2013)

Identifiants

Citer

Jin-Ying Zhang, G. Han, S.M. Chen, Y. Qian, Wei-Jiang Xu, et al.. Study of rough surface to decrease reverberation noise in ultrasonic imaging. 10th IEEE International New Circuits and Systems Conference, NEWCAS 2012, 2012, Montréal, QC, Canada. pp.509-512, ⟨10.1109/NEWCAS.2012.6329068⟩. ⟨hal-00803090⟩