Heat flow metric policy for dynamic monitoring of fixed-point cells - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Journal of Physics: Conference Series Année : 2010

Heat flow metric policy for dynamic monitoring of fixed-point cells

Résumé

In this paper, an original configuration, named as cell-within-cell, is described together with a specific measurement policy. To this end, concomitant temperature and heat flow assessments are carried out so as to obtain the exhaustive energy monitoring within the thermometric fixed-point cell. Two different melting schemes are investigated so as to consider an optimized control strategy in the near future.
Fichier principal
Vignette du fichier
G_Failleau_2010_J._Phys. _Conf._Ser._238_012012.pdf (1.96 Mo) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte

Dates et versions

hal-00661655 , version 1 (28-04-2022)

Licence

Paternité

Identifiants

Citer

Guillaume Failleau, Etienne Gaviot, Lionel Camberlein, N. Fleurence, Frédéric Polet, et al.. Heat flow metric policy for dynamic monitoring of fixed-point cells. Journal of Physics: Conference Series, 2010, 238, pp.012012/1-5. ⟨10.1088/1742-6596/238/1/012012⟩. ⟨hal-00661655⟩
121 Consultations
13 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More