Assesment of response time of SAR probe - Archive ouverte HAL
Communication Dans Un Congrès Année : 2011

Assesment of response time of SAR probe

Résumé

In this paper, the assessment of SAR probe and the prediction of SAR system measurement are presented. Usually, SAR probe have got high output impedance. Then, the response time is directly affected by the acquisition system (input impedance, cable length). Firstly a methodology is presented to define a circuit model of the SAR probe by comparison measurement and simulation. Then, it is possible to predict the time response of a SAR system including cable, matching circuit and data acquisition.
Fichier principal
Vignette du fichier
GALLEE_EBEA2011.pdf (163.07 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-00623230 , version 1 (13-09-2011)

Identifiants

  • HAL Id : hal-00623230 , version 1

Citer

François Gallée, Gaetan Guevel, Christian Person. Assesment of response time of SAR probe. EBEA 11: 10th International Congress of the European Bioelectromagnetic, Feb 2011, Rome, Italy. ⟨hal-00623230⟩
85 Consultations
102 Téléchargements

Partager

More