Analysis and artifact correction for volume correlation measurements using tomographic images from a laboratory X-ray source - Archive ouverte HAL
Article Dans Une Revue Experimental Mechanics Année : 2011

Analysis and artifact correction for volume correlation measurements using tomographic images from a laboratory X-ray source

Résumé

The effects of three artifacts (reconstruction, beam hardening and temperature of the X-ray tube) associated with the use of a lab tomograph are analyzed in terms of their induced biases for Digital Volume Correlation (DVC) from a series of reconstructed volumes acquired successively. The most detrimental effect is due to spurious dilatational strains induced by temperature variations in the tomograph. If they are not accounted for, any quantitative kinematic measurement is impossible for strain levels below 0.5 %.
Fichier principal
Vignette du fichier
EXME7-ccsd.pdf (657.32 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-00521190 , version 1 (26-09-2010)

Identifiants

Citer

Nathalie Limodin, Julien Réthoré, Jérôme Adrien, Jean-Yves Buffiere, François Hild, et al.. Analysis and artifact correction for volume correlation measurements using tomographic images from a laboratory X-ray source. Experimental Mechanics, 2011, 51 (6), pp.959-970. ⟨10.1007/s11340-010-9397-4⟩. ⟨hal-00521190⟩
221 Consultations
428 Téléchargements

Altmetric

Partager

More