Magnetic Detection Structure for Lab-on-Chip Applications Based on the Frequency Mixing Technique - Laboratoire d'Electronique et Electromagnétisme Accéder directement au contenu
Article Dans Une Revue Sensors Année : 2018

Magnetic Detection Structure for Lab-on-Chip Applications Based on the Frequency Mixing Technique

Résumé

A magnetic frequency mixing technique with a set of miniaturized planar coils was investigated for use with a completely integrated Lab-on-Chip (LoC) pathogen sensing system. The system allows the detection and quantification of superparamagnetic beads. Additionally, in terms of magnetic nanoparticle characterization ability, the system can be used for immunoassays using the beads as markers. Analytical calculations and simulations for both excitation and pick-up coils are presented; the goal was to investigate the miniaturization of simple and cost-effective planar spiral coils. Following these calculations, a Printed Circuit Board (PCB) prototype was designed, manufactured, and tested for limit of detection, linear response, and validation of theoretical concepts. Using the magnetic frequency mixing technique, a limit of detection of 15 µg/mL of 20 nm core-sized nanoparticles was achieved without any shielding.

Dates et versions

hal-01808193 , version 1 (05-06-2018)

Licence

Paternité

Identifiants

Citer

Amine Rabehi, Benjamin Garlan, Stefan Achtsnicht, Hans-Joachim Krause, Andreas Offenhäusser, et al.. Magnetic Detection Structure for Lab-on-Chip Applications Based on the Frequency Mixing Technique. Sensors, 2018, 18 (6), pp.1747. ⟨10.3390/s18061747⟩. ⟨hal-01808193⟩
329 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More